Cropico DO500CS
Calibration Standard for the DO5000 digital microhmmeter
- Upto 10A calibration current
- Switch selectable values
- Four terminal zero
The calibration Standard DO5000-CS is designed to enable the full calibration of the DO5000 digital microhmmeter. The resistance ranges are adjusted to a nominal accuracy and the actual value is measured and recorded on the calibration certificate. The value on the calibration certificate is the value that should be used when calibrating the DO5000.
| Measurement | 4 wire eliminates lead resistance |
| Display | LCD Graphics panel with backlight 30,000 count |
| Ranges | Manual / Automatic |
| Auto Zero | Value Average Hold |
| Systems Interface | RS232 fitted as standard, IEEE-488, PLC option |
| Terminals | 4mm safety sockets |
| Working Temperature | 0°C ... 45°C |
| Storage Temperature | -10°C...+60 °C |
| Mains Supply | 230V ±10%, 115 V ±10% 47...63Hz |
| Safety | IEC 1010 Class 1 |
| Dimensions | 339 x 324 x 131 mm (W D H) |
| Weight | 8.5kg |
| Accessories | Supplied with RS232 interface, mains cord, handbook and certificate of conformity |
| Range | Resolution | Min. Current | Max. Current | Accuracy at full rated current |
| 30kOhm | 1Ohm | 10mA | 100mA | ±(0.03% Rdg +0.02% FS) |
| 3kOhm | 100mOhm | 100mA | 1mA | ±(0.03% Rdg +0.01% FS) |
| 300Ohm | 10mOhm | 1mA | 10mA | ±(0.03% Rdg +0.01% FS) |
| 30Ohm | 1mOhm | 10mA | 100mA | ±(0.03% Rdg +0.01% FS) |
| 3Ohm |