Careers and job opportunities
Share:
Q

With reference to Earth Continuity Testing, does the Code Of Practice say that 25 Amps is the preferred test current?

A

No, The IEE code of practice recommends that the user chooses EITHER a low current, 20 - 200mA, (soft) test whilst flexing the cable OR a high current, 1.5 x fuse rating up to 26 Amps (hard) test.

 

View FAQs by product

 

If you are unable to find the answer to your question on our webiste, please email support@seaward.co.uk and we will get back to you as soon as we can.

Please note, our support opening times are 08:25 - 16:55 (GMT) Mon - Thurs, and 08:25 - 14:55 (GMT) on Fri.

 
Share:

Can't find the answer your are looking for? Why not as like minded professionals the question in one of the forums or groups below?

 
Stay up to date with the latest news from Seaward
 
When joining, you agree to Seaward contacting you via email about industry news, useful guides, webinars and workshops, and Seaward products which we feel will be of interest to you. Please refer to our privacy policy for more information on how we use your details.